Lucene search

K
cve[email protected]CVE-2018-14781
HistoryAug 13, 2018 - 9:48 p.m.

CVE-2018-14781

2018-08-1321:48:01
CWE-287
CWE-294
web.nvd.nist.gov
28
cve-2018-14781
medtronic
mmt 508
mmt 722
mmt 723
mmt 723k
mmt 751
paradigm real-time
paradigm revel
insulin pump
capture-replay attack
wireless vulnerability

2.9 Low

CVSS2

Attack Vector

ADJACENT_NETWORK

Attack Complexity

MEDIUM

Authentication

NONE

Confidentiality Impact

NONE

Integrity Impact

PARTIAL

Availability Impact

NONE

AV:A/AC:M/Au:N/C:N/I:P/A:N

5.3 Medium

CVSS3

Attack Vector

ADJACENT

Attack Complexity

HIGH

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

NONE

Integrity Impact

HIGH

Availability Impact

NONE

CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N

5.3 Medium

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

30.7%

Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

Affected configurations

NVD
Node
medtronicdiabetes508_minimed_insulin_pump_firmwareMatch-
AND
medtronicdiabetes508_minimed_insulin_pumpMatch-
Node
medtronicdiabetes522_paradigm_real-time_firmwareMatch-
AND
medtronicdiabetes522_paradigm_real-timeMatch-
Node
medtronicdiabetes722_paradigm_real-time_firmwareMatch-
AND
medtronicdiabetes722_paradigm_real-timeMatch-
Node
medtronicdiabetes523_paradigm_revel_firmwareMatch-
AND
medtronicdiabetes523_paradigm_revelMatch-
Node
medtronicdiabetes723_paradigm_revel_firmwareMatch-
AND
medtronicdiabetes723_paradigm_revelMatch-
Node
medtronicdiabetes523k_paradigm_revel_firmwareMatch-
AND
medtronicdiabetes523k_paradigm_revelMatch-
Node
medtronicdiabetes723k_paradigm_revel_firmwareMatch-
AND
medtronicdiabetes723k_paradigm_revelMatch-
Node
medtronicdiabetes551_minimed_530g_firmwareMatch-
AND
medtronicdiabetes551_minimed_530gMatch-
Node
medtronicdiabetes751_minimed_530g_firmwareMatch-
AND
medtronicdiabetes751_minimed_530gMatch-

CNA Affected

[
  {
    "product": "Medtronic insulin pump",
    "vendor": "ICS-CERT",
    "versions": [
      {
        "status": "affected",
        "version": "MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G"
      }
    ]
  }
]

2.9 Low

CVSS2

Attack Vector

ADJACENT_NETWORK

Attack Complexity

MEDIUM

Authentication

NONE

Confidentiality Impact

NONE

Integrity Impact

PARTIAL

Availability Impact

NONE

AV:A/AC:M/Au:N/C:N/I:P/A:N

5.3 Medium

CVSS3

Attack Vector

ADJACENT

Attack Complexity

HIGH

Privileges Required

NONE

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

NONE

Integrity Impact

HIGH

Availability Impact

NONE

CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N

5.3 Medium

AI Score

Confidence

High

0.001 Low

EPSS

Percentile

30.7%

Related for CVE-2018-14781