CVSS2
Attack Vector
NETWORK
Attack Complexity
LOW
Authentication
NONE
Confidentiality Impact
COMPLETE
Integrity Impact
COMPLETE
Availability Impact
COMPLETE
AV:N/AC:L/Au:N/C:C/I:C/A:C
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
AI Score
Confidence
High
EPSS
Percentile
54.1%
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 with Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 with Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions with Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.
Vendor | Product | Version | CPE |
---|---|---|---|
siemens | rapidpoint_400_firmware | - | cpe:2.3:o:siemens:rapidpoint_400_firmware:-:*:*:*:*:*:*:* |
siemens | rapidpoint_400 | - | cpe:2.3:h:siemens:rapidpoint_400:-:*:*:*:*:*:*:* |
siemens | rapidpoint_500_firmware | * | cpe:2.3:o:siemens:rapidpoint_500_firmware:*:*:*:*:*:*:*:* |
siemens | rapidpoint_500 | - | cpe:2.3:h:siemens:rapidpoint_500:-:*:*:*:*:*:*:* |
siemens | rapidlab_1200_firmware | * | cpe:2.3:o:siemens:rapidlab_1200_firmware:*:*:*:*:*:*:*:* |
siemens | rapidlab_1200 | - | cpe:2.3:h:siemens:rapidlab_1200:-:*:*:*:*:*:*:* |
[
{
"product": "RAPIDLab 1200 systems RAPIDPoint 400 systems RAPIDPoint 500 systems RAPIDLab 1200 Series RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 500 systems RAPIDPoint 400 systems",
"vendor": "Siemens AG",
"versions": [
{
"status": "affected",
"version": "RAPIDLab 1200 systems RAPIDPoint 400 systems / RAPIDPoint 500 systems All versions _without_ use of Siemens Healthineers Informatics products RAPIDLab 1200 Series All versions < V3.3 _with_Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions >= V3.0 _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems V2.4.X _with_ Siemens Healthineers Informatics products RAPIDPoint 500 systems All versions =< V2.3 _with_ Siemens Healthineers Informatics products RAPIDPoint 400 systems All versions_with_ Siemens Healthineers Informatics products"
}
]
}
]
CVSS2
Attack Vector
NETWORK
Attack Complexity
LOW
Authentication
NONE
Confidentiality Impact
COMPLETE
Integrity Impact
COMPLETE
Availability Impact
COMPLETE
AV:N/AC:L/Au:N/C:C/I:C/A:C
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
AI Score
Confidence
High
EPSS
Percentile
54.1%