The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network.
[
{
"product": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor",
"vendor": "Contec Health",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
]