CVSS2
Attack Vector
NETWORK
Attack Complexity
MEDIUM
Authentication
NONE
Confidentiality Impact
PARTIAL
Integrity Impact
PARTIAL
Availability Impact
NONE
AV:N/AC:M/Au:N/C:P/I:P/A:N
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
REQUIRED
Scope
CHANGED
Confidentiality Impact
LOW
Integrity Impact
LOW
Availability Impact
NONE
CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
EPSS
Percentile
42.3%
This advisory was originally posted to the NCCIC Portal on March 23, 2017, and is being released to the ICS-CERT web site.
Marc Ruef and Rocco Gagliardi of scip AG have identified an open redirect vulnerability in B. Braun Medical’s SpaceCom module, which is integrated into the SpaceStation docking station. B. Braun has produced a software update for the SpaceCom module to mitigate this vulnerability.
This vulnerability could be exploited remotely.
The following versions of the SpaceCom module, used with the SpaceStation docking station, are affected:
Successful exploitation of this vulnerability may allow URL redirection to untrusted web sites.
B. Braun Medical Inc. is a US-based company and is part of a global B. Braun group of companies.
The affected product, SpaceCom module, is a data communication interface that is used with a SpaceStation docking station for multiple intravenous infusion pumps. According to B. Braun Medical, SpaceCom modules are deployed across the Healthcare and Public Health Sector. B. Braun Medical estimates that this product is only used in the United States.
The web server of the affected product accepts untrusted input which could allow attackers to redirect the request to an unintended URL contained within untrusted input.
CVE-2017-6018NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2017-6018, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory. has been assigned to this vulnerability. A CVSS v3 base score of 5.4 has been assigned; the CVSS vector string is (AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N).CVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:U/C:L/I:L/A:N, web site last accessed March 23, 2017.
This vulnerability could be exploited remotely.
No known public exploits specifically target this vulnerability.
An attacker with a low skill level would be able to exploit this vulnerability.
B. Braun Medical has released a software update, Version 012U000040, for the SpaceCom module, which is a data communication interface used with the SpaceStation docking station. B. Braun Medical’s software update, Version 012U000040, along with instructions for downloading and installing the software update, are available on the B. Braun Medical portal at the following location (with valid account):
B. Braun Medical has directly notified affected customers using the SpaceCom module to inform them of the software update.
For B. Braun Medical portal account assistance, contact B. Braun Technical Support at:
1-800-627-7867
or
ICS-CERT in collaboration with B. Braun Medical recommends that users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:
ICS-CERT also provides a section for security recommended practices on the ICS-CERT web page at http://ics-cert.us-cert.gov/content/recommended-practices. ICS-CERT reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the ICS‑CERT Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT web site (http://ics-cert.us-cert.gov/).
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to ICS-CERT for tracking and correlation against other incidents.
cisasurvey.gov1.qualtrics.com/jfe/form/SV_9n4TtB8uttUPaM6?product=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-082-02
ftp.bbmus.com
public.govdelivery.com/accounts/USDHSCISA/subscriber/new?topic_id=USDHSCISA_138
twitter.com/CISAgov
twitter.com/intent/tweet?text=B.%20Braun%20Medical%20SpaceCom%20Open%20Redirect%20Vulnerability+https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-082-02
www.dhs.gov
www.dhs.gov/foia
www.dhs.gov/performance-financial-reports
www.facebook.com/CISA
www.facebook.com/sharer/sharer.php?u=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-082-02&title=B.%20Braun%20Medical%20SpaceCom%20Open%20Redirect%20Vulnerability
www.instagram.com/cisagov
www.linkedin.com/company/cybersecurity-and-infrastructure-security-agency
www.linkedin.com/sharing/share-offsite/?url=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-082-02
www.oig.dhs.gov/
www.usa.gov/
www.whitehouse.gov/
www.youtube.com/@cisagov
mailto:?subject=B.%20Braun%20Medical%20SpaceCom%20Open%20Redirect%20Vulnerability&body=www.cisa.gov/news-events/ics-medical-advisories/icsma-17-082-02
CVSS2
Attack Vector
NETWORK
Attack Complexity
MEDIUM
Authentication
NONE
Confidentiality Impact
PARTIAL
Integrity Impact
PARTIAL
Availability Impact
NONE
AV:N/AC:M/Au:N/C:P/I:P/A:N
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
REQUIRED
Scope
CHANGED
Confidentiality Impact
LOW
Integrity Impact
LOW
Availability Impact
NONE
CVSS:3.0/AV:N/AC:L/PR:N/UI:R/S:C/C:L/I:L/A:N
EPSS
Percentile
42.3%