MedSec has identified an improper input validation vulnerability in BMC Medical’s and 3B Medical’s Luna continuous positive airway pressure (CPAP) therapy machine. For devices released after July 1, 2017, this vulnerability has been addressed. For devices released prior to July 1, 2017, BMC Medical and 3B Medical offer no mitigations.
The following versions of the Luna CPAP Machine are affected:
Successful exploitation of this vulnerability could allow an attacker to cause a crash of the device’s Wi-Fi module resulting in a denial-of-service condition affecting the Wi-Fi module chipset. This does not affect the device’s ability to deliver therapy.
The operation of the Luna CPAP can be broken into therapeutic and communication functions. BMC Medical, based in China, manufactures the device and firmware chipset used in delivery of therapy. 3B Medical, based in the United States, manages the Wi-Fi module chipset used in communication.
The affected product, the Luna CPAP Machine, is a continuous positive airway pressure therapy machine. It is deployed worldwide across the Healthcare and Public Health sector.
An improper input validation vulnerability has been identified, which may allow an authenticated attacker to crash the CPAP’s Wi-Fi module resulting in a denial-of-service condition.
Note that the vulnerability affects only the Wi-Fi module; the device can continue delivering therapy even after the Wi-Fi module has crashed.
CVE-2017-12701NVD, http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2017-12701, NIST uses this advisory to create the CVE web site report. This web site will be active sometime after publication of this advisory. has been assigned to this vulnerability. A CVSS v3 base score of 4.6 and a has been assigned; the CVSS vector string is (AV:A/AC:L/PR:L/UI:N/S:U/C:N/I:L/A:L).CVSS Calculator, https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:L/PR:L/UI:N/S:U/C:N/I:L/A:L, web site last accessed August 15, 2017.
This vulnerability is exploitable via adjacent network access.
No known public exploits specifically target this vulnerability.
An attacker with a low skill would be able to exploit this vulnerability.
This vulnerability has been addressed in devices released after July 1, 2017. For devices released prior to July 1, 2017, BMC Medical and 3B Medical offer no mitigations.
NCCIC/ICS-CERT recommends that users take defensive measures to minimize the risk of exploitation of this vulnerability. Specifically, users should:
ICS-CERT also provides a section for security recommended practices on the ICS-CERT web page at http://ics-cert.us-cert.gov/content/recommended-practices. ICS-CERT reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly available in the ICS‑CERT Technical Information Paper, ICS-TIP-12-146-01B–Targeted Cyber Intrusion Detection and Mitigation Strategies, that is available for download from the ICS-CERT web site (http://ics-cert.us-cert.gov/).
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to ICS-CERT for tracking and correlation against other incidents.
cisasurvey.gov1.qualtrics.com/jfe/form/SV_9n4TtB8uttUPaM6?product=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-227-01
public.govdelivery.com/accounts/USDHSCISA/subscriber/new?topic_id=USDHSCISA_138
twitter.com/CISAgov
twitter.com/intent/tweet?text=BMC%20Medical%20and%203B%20Medical%20Luna%20CPAP%20Machine+https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-227-01
www.dhs.gov
www.dhs.gov/foia
www.dhs.gov/performance-financial-reports
www.facebook.com/CISA
www.facebook.com/sharer/sharer.php?u=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-227-01&title=BMC%20Medical%20and%203B%20Medical%20Luna%20CPAP%20Machine
www.instagram.com/cisagov
www.linkedin.com/company/cybersecurity-and-infrastructure-security-agency
www.linkedin.com/sharing/share-offsite/?url=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-17-227-01
www.oig.dhs.gov/
www.usa.gov/
www.whitehouse.gov/
www.youtube.com/@cisagov
mailto:?subject=BMC%20Medical%20and%203B%20Medical%20Luna%20CPAP%20Machine&body=www.cisa.gov/news-events/ics-medical-advisories/icsma-17-227-01
CVSS2
Attack Vector
NETWORK
Attack Complexity
LOW
Authentication
SINGLE
Confidentiality Impact
NONE
Integrity Impact
NONE
Availability Impact
PARTIAL
AV:N/AC:L/Au:S/C:N/I:N/A:P
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
LOW
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
NONE
Integrity Impact
NONE
Availability Impact
HIGH
CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:U/C:N/I:N/A:H
EPSS
Percentile
30.5%