ATTENTION: Exploitable remotely/low skill level to exploit
Vendor: Fujifilm
Equipment: FCR Capsula X/Carbon X
Vulnerabilities: Uncontrolled Resource Consumption, Improper Access Control
Successful exploitation of these vulnerabilities could result in a denial-of-service condition in affected cassette reader units, causing potential image loss or device unavailability. Attackers could gain unauthorized access to the underlying operating system, allowing arbitrary code execution.
The following models and versions of Fujifilm Computed Radiography cassette readers are affected:
The device is susceptible to a denial-of-service condition as a result of an overflow of TCP packets, which requires the device to be manually rebooted.
CVE-2019-10948 has been assigned to this vulnerability. A CVSS v3 base score of 7.5 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H).
The device provides insecure telnet services that lack authentication requirements. An attacker who successfully exploits this vulnerability may be able to access the underlying operating system.
CVE-2019-10950 has been assigned to this vulnerability. A CVSS v3 base score of 9.8 has been calculated; the CVSS vector string is (AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H).
Marc Ruef and Rocco Gagliardi of Scip AG reported these vulnerabilities to NCCIC.
Fujifilm has stated the CR-IR 357 system can be configured with what they call Secure Host functionality. This configuration of the software instructs CR-IR 357 to ignore all network traffic other than from the IP address of the Fujifilm image acquisition console. However, this configuration prevents more than one image acquisition console to share the CR-IR 357 Reader Unit. If the user has not implemented Reader Unit sharing, they may contact Fujifilm to request Secure Host functionality be enabled. If the user has implemented Reader Unit sharing, they should contact Fujifilm to discuss available options. Fujifilm can be contacted at 888-FUJI-MED (888-385-4633). Users outside the United States should contact their Fujifilm contact.
Fujifilm recommends that the first line of defense should be a compensating control of securing the userβs network. Measures should be taken to ensure only authorized devices and personnel have access to the network. Public or guest networks should be segmented, or users should use a VLAN to segregate public traffic from the private network. Administrative and technical controls should also be implemented.
NCCIC recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
<https://www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm>
NCCIC reminds organizations to perform proper impact analysis and risk assessment prior to deploying defensive measures.
NCCIC also provides a section for control systems security recommended practices on the ICS-CERT web page. Several recommended practices are available for reading and download, including Improving Industrial Control Systems Cybersecurity with Defense-in-Depth Strategies.
Additional mitigation guidance and recommended practices are publicly available on the ICS-CERT website in the Technical Information Paper, ICS-TIP-12-146-01BβTargeted Cyber Intrusion Detection and Mitigation Strategies.
Organizations observing any suspected malicious activity should follow their established internal procedures and report their findings to NCCIC for tracking and correlation against other incidents.
No known public exploits specifically target these vulnerabilities.
web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-10948
web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2019-10950
cisasurvey.gov1.qualtrics.com/jfe/form/SV_9n4TtB8uttUPaM6?product=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01
cwe.mitre.org/data/definitions/284.html
cwe.mitre.org/data/definitions/400.html
public.govdelivery.com/accounts/USDHSCISA/subscriber/new?topic_id=USDHSCISA_138
twitter.com/CISAgov
twitter.com/intent/tweet?text=Fujifilm%20FCR%20Capsula%20X/Carbon%20X+https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01
www.dhs.gov
www.dhs.gov/foia
www.dhs.gov/performance-financial-reports
www.facebook.com/CISA
www.facebook.com/sharer/sharer.php?u=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01&title=Fujifilm%20FCR%20Capsula%20X/Carbon%20X
www.fda.gov/MedicalDevices/DigitalHealth/ucm373213.htm
www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
www.instagram.com/cisagov
www.linkedin.com/company/cybersecurity-and-infrastructure-security-agency
www.linkedin.com/sharing/share-offsite/?url=https://www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01
www.oig.dhs.gov/
www.usa.gov/
www.whitehouse.gov/
www.youtube.com/@cisagov
mailto:?subject=Fujifilm%20FCR%20Capsula%20X/Carbon%20X&body=www.cisa.gov/news-events/ics-medical-advisories/icsma-19-113-01
CVSS2
Attack Vector
NETWORK
Attack Complexity
LOW
Authentication
NONE
Confidentiality Impact
COMPLETE
Integrity Impact
COMPLETE
Availability Impact
COMPLETE
AV:N/AC:L/Au:N/C:C/I:C/A:C
CVSS3
Attack Vector
NETWORK
Attack Complexity
LOW
Privileges Required
NONE
User Interaction
NONE
Scope
UNCHANGED
Confidentiality Impact
HIGH
Integrity Impact
HIGH
Availability Impact
HIGH
CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
AI Score
Confidence
High
EPSS
Percentile
58.7%