nvd[email protected]NVD:CVE-2023-0888
HistoryMar 13, 2023 - 9:15 a.m.

CVE-2023-0888

2023-03-1309:15:10
CWE-95
CWE-94
web.nvd.nist.gov
3
cve-2023-0888
wifi battery
administrative access

CVSS3

7.2

Attack Vector

NETWORK

Attack Complexity

LOW

Privileges Required

HIGH

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

AI Score

6.1

Confidence

High

EPSS

0.001

Percentile

35.4%

An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks

Affected configurations

Nvd
Node
bbraunbattery-pack_sp_with_wifi_firmwareRange053l000092global
OR
bbraunbattery-pack_sp_with_wifi_firmwareRange054u000092us
AND
bbraunbattery-pack_sp_with_wifiMatch-
VendorProductVersionCPE
bbraunbattery-pack_sp_with_wifi_firmware*cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:global:*:*:*
bbraunbattery-pack_sp_with_wifi_firmware*cpe:2.3:o:bbraun:battery-pack_sp_with_wifi_firmware:*:*:*:*:us:*:*:*
bbraunbattery-pack_sp_with_wifi-cpe:2.3:h:bbraun:battery-pack_sp_with_wifi:-:*:*:*:*:*:*:*

CVSS3

7.2

Attack Vector

NETWORK

Attack Complexity

LOW

Privileges Required

HIGH

User Interaction

NONE

Scope

UNCHANGED

Confidentiality Impact

HIGH

Integrity Impact

HIGH

Availability Impact

HIGH

CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H

AI Score

6.1

Confidence

High

EPSS

0.001

Percentile

35.4%

Related for NVD:CVE-2023-0888