Lucene search

K

Baxter Security Vulnerabilities

cve
cve

CVE-2022-26392

The Baxter Spectrum WBM (v16, v16D38) and Baxter Spectrum WBM (v17, v17D19, v20D29 to v20D32) when in superuser mode is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive...

6.5CVSS

6.7AI Score

0.001EPSS

2022-09-09 03:15 PM
32
7
cve
cve

CVE-2022-26393

The Baxter Spectrum WBM is susceptible to format string attacks via application messaging. An attacker could use this to read memory in the WBM to access sensitive information or cause a Denial of Service (DoS) on the...

8.1CVSS

7.7AI Score

0.001EPSS

2022-09-09 03:15 PM
35
9
cve
cve

CVE-2022-26394

The Baxter Spectrum WBM does not perform mutual authentication with the gateway server host. This may allow an attacker to perform a man in the middle attack that modifies parameters making the network connection...

5.5CVSS

6.2AI Score

0.0005EPSS

2022-09-09 03:15 PM
22
7
cve
cve

CVE-2022-26390

The Baxter Spectrum Wireless Battery Module (WBM) stores network credentials and PHI (only applicable to Spectrum IQ pumps using auto programming) in unencrypted form. An attacker with physical access to a device that hasn't had all data and settings erased may be able to extract sensitive...

4.2CVSS

5.1AI Score

0.001EPSS

2022-09-09 03:15 PM
28
6
cve
cve

CVE-2020-12047

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24), when used with a Baxter Spectrum v8.x (model 35700BAX2) in a factory-default wireless configuration enables an FTP service with hard-coded...

9.8CVSS

9.4AI Score

0.002EPSS

2020-06-29 02:15 PM
25
cve
cve

CVE-2020-12048

Phoenix Hemodialysis Delivery System SW 3.36 and 3.40, The Phoenix Hemodialysis device does not support data-in-transit encryption (e.g., TLS/SSL) when transmitting treatment and prescription data on the network between the Phoenix system and the Exalis dialysis data management tool. An attacker...

7.5CVSS

7.4AI Score

0.001EPSS

2020-06-29 02:15 PM
22
cve
cve

CVE-2020-12039

Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration...

2.4CVSS

4.1AI Score

0.001EPSS

2020-06-29 02:15 PM
23
cve
cve

CVE-2020-12040

Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that...

9.8CVSS

9.3AI Score

0.002EPSS

2020-06-29 02:15 PM
30
cve
cve

CVE-2020-12037

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could...

7.5CVSS

7.4AI Score

0.001EPSS

2020-06-29 02:15 PM
29
cve
cve

CVE-2020-12041

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) telnet Command-Line Interface, grants access to sensitive data stored on the WBM that permits temporary configuration changes to network settings of the WBM, and allows the WBM to be rebooted. Temporary configuration changes to...

9.4CVSS

9.1AI Score

0.002EPSS

2020-06-29 02:15 PM
24
cve
cve

CVE-2020-12016

Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 have hard-coded administrative account...

9.8CVSS

9.3AI Score

0.002EPSS

2020-06-29 02:15 PM
23
cve
cve

CVE-2020-12032

Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems store device data with sensitive information in an unencrypted database. This could allow an attacker with network access to view or modify sensitive data including...

9.1CVSS

8.7AI Score

0.002EPSS

2020-06-29 02:15 PM
28
cve
cve

CVE-2020-12043

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when configured for wireless networking the FTP service operating on the WBM remains operational until the WBM is...

9.8CVSS

9.5AI Score

0.002EPSS

2020-06-29 02:15 PM
23
cve
cve

CVE-2020-12012

Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials...

6.1CVSS

6.2AI Score

0.001EPSS

2020-06-29 02:15 PM
22
cve
cve

CVE-2020-12035

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings...

4.9CVSS

5.1AI Score

0.0004EPSS

2020-06-29 02:15 PM
27
cve
cve

CVE-2020-12036

Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could...

7.5CVSS

7.4AI Score

0.001EPSS

2020-06-29 02:15 PM
26
cve
cve

CVE-2020-12020

Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13 and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 does not restrict non administrative users from gaining access to the operating system and editing the application startup script. Successful exploitation of this vulnerability may allow an...

6.1CVSS

6.2AI Score

0.0004EPSS

2020-06-29 02:15 PM
21
cve
cve

CVE-2020-12024

Baxter ExactaMix EM 2400 versions 1.10, 1.11, 1.13, 1.14 and ExactaMix EM1200 Versions 1.1, 1.2, 1.4 and 1.5 does not restrict access to the USB interface from an unauthorized user with physical access. Successful exploitation of this vulnerability may allow an attacker with physical access to the....

6.1CVSS

5.9AI Score

0.001EPSS

2020-06-29 02:15 PM
25
cve
cve

CVE-2020-12045

The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded...

9.8CVSS

9.4AI Score

0.002EPSS

2020-06-29 02:15 PM
22
cve
cve

CVE-2020-12008

Baxter ExactaMix EM 2400 Versions 1.10, 1.11 and ExactaMix EM1200 Versions 1.1, 1.2 systems use cleartext messages to communicate order information with an order entry system. This could allow an attacker with network access to view sensitive data including...

7.5CVSS

7.2AI Score

0.002EPSS

2020-06-29 02:15 PM
20
cve
cve

CVE-2000-1151

Baxter IRC client in BeOS r5 pro and earlier allows remote attackers to conduct a denial of service via a message that contains a long...

7AI Score

0.003EPSS

2001-01-09 05:00 AM
26